SprayQuantAI® Technical Specifications

SprayQuantAI® combines an optical measurement probe with a control unit and evaluation software. The probe records time-resolved light-scattering signals from droplets. The control unit acquires and processes these signals and provides measurement results to the host computer or external systems.

Measurement Probe

The optical probe defines the measurement region and detects the scattered light generated by droplets passing through it. The optical configuration is selected according to the spray, installation space and required working distance.

ParameterSpecification
Measurement objectDroplets in sprays and flows
Measurement methodTime-resolved light-scattering signal analysis
Typical working distance125 mm
Available working distances62.5 mm, 125 mm and 250 mm
Typical laser wavelength405 nm
Typical laser output power3 mW
Intended minimum droplet sizeApproximately 1 µm
Probe dimensions182 × 74.9 × 32 mm

The wavelength and laser power shown above describe the usual SprayQuantAI® configuration. Other optical configurations may be selected for specific measurement tasks.

Control Unit

The control unit acquires the optical detector signals, runs the configured calculation electronics and communicates with the host computer. It also provides interfaces for connecting SprayQuantAI® to test equipment and process-control systems.

ParameterSpecification
Host communicationEthernet
Digital inputs4
Digital outputs4
Current-loop inputs4
Current-loop outputs4
Trigger interfacesTrigger input and trigger output
Dimensions251 × 211 × 109 mm

The available analog and digital interfaces depend on the control-unit configuration. Confirm the required inputs and outputs when specifying a system for integration.

Measurement Results

In its standard measurement mode, SprayQuantAI® determines droplet size, velocity and number or rate. Additional results depend on the selected evaluation method, system configuration and measurement task.

System Configurations

SystemOptical probeControl unit
SQA1LSS1ZEON
SQA2LSS2IMEA

The final system configuration depends on the measurement task and required interfaces. Contact ai-quanton to discuss the optical setup and integration requirements for your application.

Contact ai-quanton